Isometric Force-Time Curve Assessment: Accuracy, Precision and Repeatability of a Mobile Application and Portable and Lightweight Device—The DINABANG®

Santos D, Bravo-Sánchez A, Peyré-Tartaruga LA, Simini F, Zacca R.

2024
Primer autor
Santos D
Cita
2024
DOI
10.20944/preprints202408.0767.v1

Resumen

Background: Strength assessment is one of the main fields in sports performance, physical rehabilitation, physical activity and health. Purpose: To compare maximal voluntary isometric contractions (MVIC) and paired voluntary isometric contractions (VIC) of knee extensors between an isokinetic dynamometer (BIODEX) and a portable and lightweight device (DINABANG). Methods: From 19 volunteers (age: 28.7±7 y; body mass: 72±10 kg and height: 173±7 cm) we obtained 114 paired MVIC measures and, from the force-time curves of these repetitions, 22507 paired VIC measures of knee extensors. Results: We observed “excellent” repeatability for MVIC (ICC:1.00; p<0.001) between BIODEX (247±79.5 Nm) and DINABANG (247±74.8 Nm), with “trivial” effect (mean difference: 0.12 Nm (0.02%); 95%CI: -0.13 to 0.23 Nm; p=0.606; d=0.048). Bland-Altman plots revealed high accuracy for MVIC (bias: 0.12 Nm) and consistent distribution (precision) inside the limits of agreement (-4.81 to 5.06 Nm) and respective 95%CI. “Excellent” repeatability was also observed for VIC (ICC:1.00; p<0.001) between BIODEX (219±84.1 Nm) and DINABANG (218±84.0 Nm), with “trivial” effect (0.24 Nm (0.11%); 0.08 to 0.11 Nm; p<0.001; d=0.100). Bland-Altman plots revealed high accuracy for VIC (bias: 0.24 Nm) and consistent distribution (precision) inside the limits of agreement (-4.5 to 4.9 Nm) and respective 95%CI. Conclusion: DINABANG is accurate, precise and reliable for torque measurement.